Daiichi Nekken TB-IIG / TB-IIS Zirconia O2 Analyzer Sensor (Used)

Item Code:

TB-IIG / TB-IIS

Brand:

DAIICHI NEKKEN

Condition:

Refurbished


Daiichi Nekken TB-IIG / TB-IIS Zirconia O2 Analyzer Sensor (Used)

Product Information

Product Name Zirconia O2 Analyzer Sensor (detector)
Manufacturer DAIICHI NEKKEN Co., Ltd., Japan
Model 1 TB-IIG (TB-2G) – Ser. No. T-11780-1101
Model 2 TB-IIS (TB-2S) – Ser. No. T-10250-0707
Series TB-II zirconia O2 sensor series – TB-IIG, TB-IIS, TB-IIF (P/R/S/V), TB-IIV, TB-IIVN available
Sensing Element Stabilized zirconia ceramic
Body Stainless steel housing with sample gas inlet/outlet fittings and top cable connector
Labels "Zirconia O2 Analyzer", "CAUTION HIGH TEMPERATURE"
Country of Manufacture Japan
Condition Used
Supplier SM MARINE SERVICE™ – worldwide shipping

Product Introduction

The Daiichi Nekken TB-IIG and TB-IIS are zirconia O2 analyzer sensors (detectors) used to measure oxygen in gas streams. On ships they are mainly used in the inert gas system (IGS) of tankers and in boiler exhaust / flue gas O2 monitoring. SM MARINE SERVICE™ (https://www.smmarineservice.com/) supplies used Daiichi Nekken TB-II sensors and related O2 analyzer spares worldwide.

What Is This Equipment?

A zirconia O2 sensor uses a stabilized zirconia ceramic element heated to high temperature. Oxygen ions move through the hot ceramic, creating a voltage that depends on the oxygen difference between the sample gas and the reference air. The sensor works together with a Daiichi Nekken control unit, which converts the signal into an O2 reading, alarms and outputs. On tankers, the inert gas system mixes nitrogen or boiler exhaust gas into the cargo tanks to prevent explosions, and Daiichi Nekken O2 meters built into the IGS measure the O2 concentration of this gas; the manufacturer reports use on more than 1,000 vessels. The TB-IIG is the exhaust/combustion gas version and can also show CO+H2 during fuel-rich combustion as negative O2 %.

Main Features

  • Stabilized zirconia ceramic detection element
  • Multi-layer protection of the zirconia sensor against corrosive gases such as HCl and SOx
  • Simple structure for easy maintenance
  • TB-IIG: combustion exhaust gas O2, range -20 to 25 % O2
  • TB-II F series: clean gas from ppm O2 up to 100 % O2
  • Stainless steel body with threaded sample connections
  • Works with Daiichi Nekken control units (C-28C, C-48 series)

Daiichi Nekken TB-IIG TB-IIS zirconia O2 analyzer sensor used - SM MARINE SERVICE
Daiichi Nekken TB-IIS (left) and TB-IIG (right) zirconia O2 analyzer sensors – SM MARINE SERVICE™ photo

Main Technical Specifications

Detection Principle Zirconia (stabilized zirconia ceramic) oxygen sensor
TB-IIG Measuring Range -20 to 25 % O2 (combustion exhaust gas)
TB-IIG Sample Temperature 300 °C max at the sensor; up to 700 °C with guide probe, up to 1350 °C with aspirator block
TB-IIG Sample Pressure 0 ±5 kPa
TB-II F Measuring Range 1 ppm to 100 % O2
TB-II F Sample Flow 0.2 to 5 L/min (depending on version)
Warm-up Time Approx. 20 minutes
Control Unit Outputs (C-28C / C-48) 4–20 mA, 0–1/5/10 V, RS232C, 2 alarm points, sensor temperature fault alarm

Inside the Unit: PCBs, Modules & Part Numbers

The TB-II sensors are available in the following models and with these related units. Send us your model and serial number and we will supply the matching sensor.

TB-IIG (TB-2G) Zirconia O2 Sensor – Ser. No. T-11780-1101

Exhaust / combustion gas O2 sensor for boilers, furnaces and IGS.

TB-IIS (TB-2S) Zirconia O2 Sensor – Ser. No. T-10250-0707

TB-II series zirconia O2 sensor, S version.

TB-IIF-P / TB-IIF-R / TB-IIF-S / TB-IIF-V

Flow-through O2 sensors for clean gas (Rc 1/8, Rc 1/4, 3/8" Swagelok and 4VCR connections).

TB-IIV / TB-IIVN

Direct-insertion O2 sensors for vacuum furnaces and sealed vessels.

Zirconia Sensing Element and Heater

The heated zirconia ceramic cell inside the stainless housing that produces the O2 signal.

Control Units C-28C, C-48, C-58, C-101

Daiichi Nekken control/display units that power the sensor heater, show the O2 value and give alarms and 4–20 mA output.

Main Components & Related Units

  • Daiichi Nekken control unit (C-28C, C-48, C-58)
  • Aspirator block and guide probe for TB-IIG sampling
  • Sample gas line, filter and calibration gas connection
  • Inert gas generator / boiler flue gas system
  • TB-FI portable zirconia O2 analyzers

Marine & Industrial Applications

  • Inert gas systems (IGS) on oil tankers, chemical tankers and product carriers
  • Inert gas generators and nitrogen generators
  • Boiler exhaust / flue gas O2 measurement and combustion control
  • Industrial furnaces, kilns, incinerators and heat treatment

Daiichi Nekken TB-II zirconia O2 sensor label TB-IIG TB-IIS marine IGS oxygen analyzer - SM MARINE SERVICE
Model labels of the TB-IIG and TB-IIS zirconia O2 sensors

Common Problems & Troubleshooting

Safety note: The sensor runs at high temperature – let it cool before handling. Isolate power at the control unit. Purge sample lines before opening them, as exhaust or inert gas has very low oxygen.

  • Reading too high – air leakage in the sample line or fittings; check joints and the sensor seal.
  • Slow response – blocked filter or sample line, or aged zirconia element.
  • Sensor temperature fault – heater or cable problem; check the connector and the control unit.
  • Unstable reading – condensation or dust in the sample gas; check the sampling system.
  • Calibration drift – recalibrate with zero/span gas or air as per the manual.

Maintenance Tips

  • Check calibration regularly with reference gas and air
  • Keep sample lines, filters and fittings clean and leak-free
  • Protect the sensor from water, oil and condensate
  • Check the cable connector at the top of the sensor
  • Allow full warm-up (about 20 minutes) before taking readings

Spare Parts & Replacement Availability

SM MARINE SERVICE™ supplies used Daiichi Nekken TB-II sensors and can source control units and related spares. For a fast quotation please send:

  • Sensor model and serial number from the label
  • Control unit model
  • Application (IGS, boiler, nitrogen generator)
  • Vessel name and delivery port

We ship worldwide.

Why This Equipment Is Still Used

Daiichi Nekken zirconia O2 meters are fitted in the inert gas systems of many tankers. A matching TB-II sensor fits the existing control unit, piping and wiring, so it is the simplest way to restore O2 measurement and keep the IGS working.

FAQ

1. What is the Daiichi Nekken TB-IIG?
A zirconia O2 sensor for combustion exhaust gas, used in boilers and inert gas systems.

2. What is the TB-IIS?
A TB-II series zirconia O2 sensor (S version) from Daiichi Nekken.

3. What is the measuring range of the TB-IIG?
-20 to 25 % O2.

4. Which control unit is used with the TB-II sensors?
Daiichi Nekken control units such as C-28C and C-48 (also C-58).

5. Is TB-2G the same as TB-IIG?
Yes, TB-2G is another way of writing TB-IIG.

6. Where is this sensor used on a ship?
Mainly in the inert gas system (IGS) of tankers and in boiler flue gas monitoring.

7. Why is my O2 reading too high?
Usually air is leaking into the sample line; check all fittings.

8. Do you have TB-IIF sensors?
Send us the full model (TB-IIF-P, -R, -S, -V) and we will check.

9. Are these sensors new?
No, they are used.

10. Do you ship worldwide?
Yes, SM MARINE SERVICE™ ships worldwide.

Short Product Description

Used Daiichi Nekken zirconia O2 analyzer sensors, model TB-IIG (Ser. No. T-11780-1101) and model TB-IIS (Ser. No. T-10250-0707), supplied by SM MARINE SERVICE™. These Japanese stainless steel sensors use a heated stabilized zirconia ceramic element to measure oxygen in gas, working with Daiichi Nekken control units such as C-28C and C-48. They are widely used in tanker inert gas systems (IGS), inert gas generators and boiler exhaust O2 monitoring. The TB-IIG covers -20 to 25 % O2 for combustion gas. Other TB-II versions and control units can also be sourced. Worldwide shipping available.

SM MARINE SERVICE™ – Marine Automation & Navigation Equipment Supplier

Available Equipment Conditions: New • Used • Refurbished • Reconditioned (Specific product condition and availability are subject to confirmation.)
Stock Locations: Bangladesh • China
Services: Worldwide Shipping • Technical Support • Spare Parts Supply • Marine Equipment Sourcing
Samir Market, DT Road, Pahartali, beside CDA Market, Chattogram, Bangladesh
Mobile / WhatsApp / Viber: +88 01818 929292
Email: info@smmarineservice.com
Website: https://www.smmarineservice.com/

Need a Daiichi Nekken TB-II O2 sensor? Send us your model and serial number today for a fast quotation.


Related products

Seres ODME MK III Oil Discharge Monitoring Equipment Control Unit

BDT 0

Seres ODME MK III Oil Discharge Monitoring Equipment Control Unit

BDT 0

Seres ODME MK III Oil Discharge Monitoring Equipment Control Unit

BDT 0

KONGSBERG GT404D4H25T Pressure Transmitter | GT404 25 Bar Marine Cargo Line & Manifold Pressure Sensor

BDT 0

SM Marine Service

SM Marine Service

Marine Equipment Supplier

Welcome to SM Marine Service! How can we help you today? 4:47 AM
Need Help? Chat with us