TB-IIG / TB-IIS
DAIICHI NEKKEN
Refurbished
| Product Name | Zirconia O2 Analyzer Sensor (detector) |
| Manufacturer | DAIICHI NEKKEN Co., Ltd., Japan |
| Model 1 | TB-IIG (TB-2G) – Ser. No. T-11780-1101 |
| Model 2 | TB-IIS (TB-2S) – Ser. No. T-10250-0707 |
| Series | TB-II zirconia O2 sensor series – TB-IIG, TB-IIS, TB-IIF (P/R/S/V), TB-IIV, TB-IIVN available |
| Sensing Element | Stabilized zirconia ceramic |
| Body | Stainless steel housing with sample gas inlet/outlet fittings and top cable connector |
| Labels | "Zirconia O2 Analyzer", "CAUTION HIGH TEMPERATURE" |
| Country of Manufacture | Japan |
| Condition | Used |
| Supplier | SM MARINE SERVICE™ – worldwide shipping |
The Daiichi Nekken TB-IIG and TB-IIS are zirconia O2 analyzer sensors (detectors) used to measure oxygen in gas streams. On ships they are mainly used in the inert gas system (IGS) of tankers and in boiler exhaust / flue gas O2 monitoring. SM MARINE SERVICE™ (https://www.smmarineservice.com/) supplies used Daiichi Nekken TB-II sensors and related O2 analyzer spares worldwide.
A zirconia O2 sensor uses a stabilized zirconia ceramic element heated to high temperature. Oxygen ions move through the hot ceramic, creating a voltage that depends on the oxygen difference between the sample gas and the reference air. The sensor works together with a Daiichi Nekken control unit, which converts the signal into an O2 reading, alarms and outputs. On tankers, the inert gas system mixes nitrogen or boiler exhaust gas into the cargo tanks to prevent explosions, and Daiichi Nekken O2 meters built into the IGS measure the O2 concentration of this gas; the manufacturer reports use on more than 1,000 vessels. The TB-IIG is the exhaust/combustion gas version and can also show CO+H2 during fuel-rich combustion as negative O2 %.

Daiichi Nekken TB-IIS (left) and TB-IIG (right) zirconia O2 analyzer sensors – SM MARINE SERVICE™ photo
| Detection Principle | Zirconia (stabilized zirconia ceramic) oxygen sensor |
| TB-IIG Measuring Range | -20 to 25 % O2 (combustion exhaust gas) |
| TB-IIG Sample Temperature | 300 °C max at the sensor; up to 700 °C with guide probe, up to 1350 °C with aspirator block |
| TB-IIG Sample Pressure | 0 ±5 kPa |
| TB-II F Measuring Range | 1 ppm to 100 % O2 |
| TB-II F Sample Flow | 0.2 to 5 L/min (depending on version) |
| Warm-up Time | Approx. 20 minutes |
| Control Unit Outputs (C-28C / C-48) | 4–20 mA, 0–1/5/10 V, RS232C, 2 alarm points, sensor temperature fault alarm |
The TB-II sensors are available in the following models and with these related units. Send us your model and serial number and we will supply the matching sensor.
Exhaust / combustion gas O2 sensor for boilers, furnaces and IGS.
TB-II series zirconia O2 sensor, S version.
Flow-through O2 sensors for clean gas (Rc 1/8, Rc 1/4, 3/8" Swagelok and 4VCR connections).
Direct-insertion O2 sensors for vacuum furnaces and sealed vessels.
The heated zirconia ceramic cell inside the stainless housing that produces the O2 signal.
Daiichi Nekken control/display units that power the sensor heater, show the O2 value and give alarms and 4–20 mA output.

Model labels of the TB-IIG and TB-IIS zirconia O2 sensors
Safety note: The sensor runs at high temperature – let it cool before handling. Isolate power at the control unit. Purge sample lines before opening them, as exhaust or inert gas has very low oxygen.
SM MARINE SERVICE™ supplies used Daiichi Nekken TB-II sensors and can source control units and related spares. For a fast quotation please send:
We ship worldwide.
Daiichi Nekken zirconia O2 meters are fitted in the inert gas systems of many tankers. A matching TB-II sensor fits the existing control unit, piping and wiring, so it is the simplest way to restore O2 measurement and keep the IGS working.
1. What is the Daiichi Nekken TB-IIG?
A zirconia O2 sensor for combustion exhaust gas, used in boilers and inert gas systems.
2. What is the TB-IIS?
A TB-II series zirconia O2 sensor (S version) from Daiichi Nekken.
3. What is the measuring range of the TB-IIG?
-20 to 25 % O2.
4. Which control unit is used with the TB-II sensors?
Daiichi Nekken control units such as C-28C and C-48 (also C-58).
5. Is TB-2G the same as TB-IIG?
Yes, TB-2G is another way of writing TB-IIG.
6. Where is this sensor used on a ship?
Mainly in the inert gas system (IGS) of tankers and in boiler flue gas monitoring.
7. Why is my O2 reading too high?
Usually air is leaking into the sample line; check all fittings.
8. Do you have TB-IIF sensors?
Send us the full model (TB-IIF-P, -R, -S, -V) and we will check.
9. Are these sensors new?
No, they are used.
10. Do you ship worldwide?
Yes, SM MARINE SERVICE™ ships worldwide.
Used Daiichi Nekken zirconia O2 analyzer sensors, model TB-IIG (Ser. No. T-11780-1101) and model TB-IIS (Ser. No. T-10250-0707), supplied by SM MARINE SERVICE™. These Japanese stainless steel sensors use a heated stabilized zirconia ceramic element to measure oxygen in gas, working with Daiichi Nekken control units such as C-28C and C-48. They are widely used in tanker inert gas systems (IGS), inert gas generators and boiler exhaust O2 monitoring. The TB-IIG covers -20 to 25 % O2 for combustion gas. Other TB-II versions and control units can also be sourced. Worldwide shipping available.
Available Equipment Conditions: New • Used • Refurbished • Reconditioned (Specific product condition and availability are subject to confirmation.)
Stock Locations: Bangladesh • China
Services: Worldwide Shipping • Technical Support • Spare Parts Supply • Marine Equipment Sourcing
Samir Market, DT Road, Pahartali, beside CDA Market, Chattogram, Bangladesh
Mobile / WhatsApp / Viber: +88 01818 929292
Email: info@smmarineservice.com
Website: https://www.smmarineservice.com/
Need a Daiichi Nekken TB-II O2 sensor? Send us your model and serial number today for a fast quotation.
Marine Equipment Supplier