Eaton MTL G1010 Mark II Oxygen Analyser Panel – Galvanic O2 Analyser (ppm) with Flowmeter

Item Code:

G1010 Mark II

Brand:

Eaton MTL (Crouse-Hinds Series)

Condition:

Refurbished


Eaton MTL G1010 Mark II Oxygen Analyser Panel – Galvanic O2 Analyser (ppm) with Flowmeter

Product Information

Analyser Manufacturer Eaton – MTL, Crouse-Hinds Series
Analyser Model G1010 MTL Oxygen analyser Mark II
Product Type Oxygen Analyser panel (wall-mounted enclosure)
Display Unit ppm (sensor cell type / range Not clearly visible in the supplied photograph)
Panel Items INFLUX flowmeter (rotameter) with needle valve ("FULL OPEN"), key switch, SELECTION CAL – SAMPLE switch, CALIBRATION SPAN – ZERO switch, four tube/cable glands
Analyser Keys / Indicators Edit, Up, Down, Cal; Alarm 1 and Alarm 2 LEDs
Panel Builder Not clearly visible in the supplied photograph
Part / Serial Number, Power Rating Not clearly visible in the supplied photograph
Condition Refurbished (subject to confirmation)
Supplier SM MARINE SERVICE™, Chattogram, Bangladesh

Product Introduction

This oxygen analyser panel is built around the Eaton MTL G1010 Mark II galvanic oxygen analyser and shows oxygen in ppm. It includes an Influx flowmeter and calibration selector switches, and is typically used to monitor oxygen in nitrogen generators and inert gas systems on ships. SM MARINE SERVICE™ (https://www.smmarineservice.com/) supplies oxygen analysers and gas analyser spares worldwide.

What Is This Equipment?

The MTL G1010 range uses a galvanic oxygen cell to measure oxygen from ppm to percent levels. On this panel the sample gas passes through the flowmeter to the sensor; the "SELECTION CAL – SAMPLE" switch selects calibration gas or sample, and the "CALIBRATION SPAN – ZERO" switch selects span or zero gas. The G1010 displays the oxygen value, has two alarm LEDs and keys for editing and calibration.

Main Features

  • Galvanic oxygen measurement, ppm display
  • Two programmable alarm contacts (Alarm 1 / Alarm 2)
  • Analogue output 0–5 V or 4–20 mA
  • Sample flowmeter with needle valve
  • CAL / SAMPLE and SPAN / ZERO selector switches
  • Complete wall-mounted panel, ready to install

Eaton MTL G1010 Mark II oxygen analyser panel ppm with Influx flowmeter - SM MARINE SERVICE
Eaton MTL G1010 Mark II oxygen analyser panel – SM MARINE SERVICE™

Main Technical Specifications

G1010 range data (Eaton catalogue):

Measurement Oxygen, galvanic cell
Cell Options / Ranges N cell 100 ppm–50%; E cell 0.1–100%; H cell 10 ppm–10%; L cell 1 ppm–10%
Outputs 0–5 V or 4–20 mA; 2 volt-free alarm contacts
Power 110–120 V / 220–240 V AC 50/60 Hz or 24 V DC
Response (T90) Approx. 3 s (% levels), 20 s (ppm levels)
Stability <2% of reading per month or 1 ppm
Sample Flow 100–300 ml/min optimum, 1000 ml/min max.
Ambient 0 to +40 °C, <80% RH
Analyser Size 96 x 96 x 196 mm panel mount, 1.0 kg
Connections 1/4" compression fittings for 6 mm tube

Inside the Unit: PCBs, Modules & Part Numbers

Eaton does not publish PCB numbers for the G1010. Board and cell part numbers are Not verified; send a photo of the rear label for exact matching.

Analyser Electronics / Display Board

Signal processing, display, alarms and outputs. PCB number: Not verified.

Galvanic Oxygen Cell

Consumable sensor cell – types N, E, H or L. Part number: Not verified.

Power Supply

AC or 24 V DC version. Part number: Not verified.

Influx Flowmeter

Sample gas rotameter with needle valve (Influx). Part number: Not clearly visible in the supplied photograph.

Main Components & Related Units

  • G1010 – integral cell version
  • G1010R – remote sensor enclosure with flowmeter
  • G1010Tx / G1010TxX – remote sensor with 4–20 mA transmitter (TxX for hazardous areas)
  • G1010RJ – Exd enclosure version
  • Calibration gas (span / zero), sample tubing and valves

Marine & Industrial Applications

  • Nitrogen generators on tankers, gas carriers and offshore units
  • Inert gas systems and tank blanketing
  • Air separation, furnaces, gloveboxes, packaging

MTL G1010 Mark II oxygen analyser display close-up Edit Cal Alarm 1 Alarm 2 ppm
Close-up of the G1010 Mark II display and keys

Common Problems & Troubleshooting

  • Reading drifts or slow response: oxygen cell ageing – calibrate or replace the cell.
  • Reading too high: air leak in sample line or fittings.
  • No flow: check flowmeter valve and sample pressure.
  • Calibration fails: check span/zero gas and switch positions.
  • Blank display: check power supply and fuse.

Safety notes: isolate power before opening the panel. Nitrogen and inert gas can cause asphyxiation – vent sample gas safely and work with gas detection in place.

Maintenance Tips

  • Calibrate regularly with certified span and zero gas.
  • Check sample tubing for leaks.
  • Replace the galvanic cell at the end of its life.

Spare Parts & Replacement Availability

SM MARINE SERVICE™ supplies MTL G1010 analysers and complete oxygen analyser panels; availability subject to confirmation. Please send the analyser rear label photo (model code, cell type, range, supply), vessel name / IMO and delivery port. We ship worldwide.

Why This Equipment Is Still Used

The G1010 is a simple, stable oxygen analyser widely fitted to nitrogen and inert gas systems; a same-type replacement keeps panel cut-outs, wiring and calibration procedures unchanged.

FAQ

1. What is the MTL G1010?
A galvanic oxygen analyser from Eaton MTL (Crouse-Hinds series).

2. What does this panel measure?
Oxygen in ppm.

3. Which cell types exist?
N, E, H and L cells for different ranges and background gases.

4. What outputs does it have?
0–5 V or 4–20 mA and two alarm contacts.

5. What do the switches do?
Select calibration or sample gas, and span or zero gas.

6. Do you supply G1010 PCBs or cells?
On request – send the rear label photo; PCB numbers are not public.

7. Where is it used on ships?
Nitrogen generators and inert gas systems.

8. Do you ship worldwide?
Yes.

Short Product Description

Oxygen analyser panel with an Eaton MTL G1010 Mark II galvanic oxygen analyser (Crouse-Hinds series), reading oxygen in ppm. The wall-mounted panel includes an Influx sample flowmeter with needle valve, a key switch, a CAL / SAMPLE selector and a SPAN / ZERO calibration selector. The G1010 offers two alarm contacts, 0–5 V or 4–20 mA output and fast response, with N, E, H and L cell options. It is typically used to monitor residual oxygen in nitrogen generators and inert gas systems on tankers and gas carriers. SM MARINE SERVICE™ supplies refurbished G1010 analysers and panels worldwide.

 


SM MARINE SERVICE™ – Marine Automation & Navigation Equipment Supplier

Available Equipment Conditions: New • Used • Refurbished • Reconditioned (Specific product condition and availability are subject to confirmation.)

Stock Locations: Bangladesh • China

Services: Worldwide Shipping • Technical Support • Spare Parts Supply • Marine Equipment Sourcing

Samir Market, DT Road, Pahartali, beside CDA Market, Chattogram, Bangladesh

Mobile / WhatsApp / Viber: +88 01818 929292

Email: info@smmarineservice.com

Website: https://www.smmarineservice.com/

 

Contact SM MARINE SERVICE™ today for a quotation on the MTL G1010 oxygen analyser and gas analyser spare parts.


BDT:

0

-
+

Related products

Deckma OMD-21 Oil Content Meter | 15 PPM Bilge Alarm

BDT 0

BRANNSTROM CLEANTRACK 1000B Oil Discharge Monitoring and Control System (ODME)

BDT 0

Saginomiya SNS-C106X60 Pressure Control Switch – Marine Refrigeration, HVAC & Compressor Protection Control

BDT 0

BD-Sensors

BDT 0

SM Marine Service

SM Marine Service

Marine Equipment Supplier

Welcome to SM Marine Service! How can we help you today? 12:23 PM
Need Help? Chat with us